Repeatability and reproducibility limits: where 2.77 comes from
Updated: September 30, 2026 · ISKRIT
Two results on the same sample never agree exactly. The repeatability limit is the largest difference you may accept before deciding something went wrong, and it is a fixed multiple of the method's standard deviation.
Calculator
Two standard deviations from the method or from your own data.
The formula and the factor
r = 2.77 · sr and R = 2.77 · sR
where sr is the repeatability standard deviation and sR the reproducibility one. The factor is not arbitrary. The difference of two independent results has a standard deviation of √2 times the standard deviation of one result, and a 95 % two-sided interval is 1.96 standard deviations wide, so the multiplier is 1.96 · √2 = 2.77.
The two sets of conditions
Repeatability conditions mean the same method, the same laboratory, the same operator, the same equipment and a short interval of time. Reproducibility conditions mean the same method on the same material in different laboratories, with different operators and equipment. Everything between those two, same laboratory but a different day or a different analyst, is intermediate precision, and using r where intermediate precision applies is the most common misuse of the limit.
Using it
Duplicate determinations giving 12.4 and 13.1 mg/l, on a method with sr = 0.28 mg/l, are compared against r = 2.77 · 0.28 = 0.78 mg/l. The difference of 0.7 is inside the limit, so the pair is accepted and the result reported as their mean. Had the difference been 0.9, the standard response is to repeat the determination, not to average anyway and not to discard the inconvenient one.
Where the standard deviations come from
They are properties of the method, established in a collaborative study or taken from the method's own precision statement, not computed from the two results being compared. A laboratory that estimates sr from the very pair it is testing has built a circular check that can never fail. Internal quality control data accumulated over months is the legitimate in-house source, and it should be revalidated when equipment or reagents change.
Questions
Where does the factor 2.77 come from? The difference of two independent results has a standard deviation √2 times that of one result, and a 95 percent two-sided interval is 1.96 standard deviations wide, so the factor is 1.96 · √2 = 2.77.
What is the difference between repeatability and reproducibility conditions? Repeatability means the same method, laboratory, operator and equipment within a short time. Reproducibility means the same method on the same material in different laboratories. Anything in between is intermediate precision, and r does not apply to it.
Repeatability and reproducibility limits, and the comparison of a result pair against them, are handled in the internal quality control tool.
Measurement uncertainty calculator: GUM budget with Type A, Type B and expanded U
Area from coordinates calculator: shoelace formula for survey plots
Bearing and distance from coordinates: survey inverse calculator
Rotating shift patterns: covering the posts and balancing the hours
Calibration intervals: setting them, shortening them, defending them
Turning an estimate into a bill of quantities that foots
Progress billing: percent complete, carry-forward and retention
Checking claimed quantities: the design schedule and the survey
Pile capacity from CPT: how the LCPC method reads a cone trace
Driven and bored piles: why the same soil gives different capacity
What belongs on a borehole log, and what quietly gets left off
Drawing a cross section: correlating layers between boreholes
From an AGS4 file to a laboratory summary: what the groups carry
Characteristic values: turning a set of test results into one number
Void ratio, porosity and saturation from one density test
When one soil layer is really two: splitting geotechnical units
ASTM D4254 minimum index density: procedure and Dr calculator
Relative density Dr and density index Id are not the same number
USCS classification step by step: ASTM D2487
Cu and Cc: what the two gradation coefficients actually tell you
Dual symbols in USCS: the 5 to 12 percent fines band
EN ISO 14688-1 against USCS: two different size ladders
AGS4 file validation: what the rules actually check
AGS4 syntax: the five rules that break a delivery
ASTM D2435 oedometer test: from the curve to Cc, mv and Eoed
Converting oedometer modulus to Young’s modulus
EN ISO 17892-5 and ASTM D2435: where the two diverge
ASTM D1196 plate load test: modulus and subgrade reaction
DIN 18134: Ev1, Ev2 and the compaction ratio
Ménard pressuremeter modulus: ASTM D4719 and EN ISO 22476-4
Hydraulic conductivity from a pumping test: Dupuit calculator and formulas
Shewhart control chart constants: A2, A3, B3, B4, D3, D4, d2, c4
The eight out of control tests on a Shewhart chart
How rebar cutting plans are optimised: the cutting stock problem
Cutting waste in rebar fabrication: what percentage is normal
Rebound hammer calibration: why the factory curve is not enough
Kerf and offcut: what a linear cutting plan has to account for
Earthwork volume from survey points: grid, TIN and cross sections
Fire load density: the formula, the calorific values and the traps
Explosive atmosphere zones 0, 1, 2 and 20, 21, 22 explained
Pollution charges per tonne: how the over-limit multiplier works
The Russian environmental fee: who pays, when, and how the advances work